FDAClass IISeptember 7, 2022

Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In recall

Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number…

Why was Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In recalled?

Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 0084276802852621340BB22120610464344840, 0084276802852622049BC23021810464344840, 0084276802852622112BF23042210464344840. Lot Numbers: lot 21340BB Exp Date: 2022-12-06; lot 22049BC (corrected lot number 9/8/2022) Exp Date: 2023-02-18; lot 22112BF Exp Date: 2023-04-20.

Product
Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
129 units US; 159 units OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Austria, France, ¿¿ Germany, Italy, Switzerland.
Recall date
September 7, 2022
Classification
Class II