FDAClass IIMay 16, 2018

IMRIS IMRI 1.5T AND 3T S System Product Usage: The recall

IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an…

Why was IMRIS IMRI 1.5T AND 3T S System Product Usage: The recalled?

The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 10111-00, 10073-01, 10043-01, 10097-01, 10006-11, 10076-21, 10081-01, 10089-01, 10101-01, 10093-01

Product
IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an…
Recalling firm
Deerfield Imaging, Inc.
Quantity
10 units
Distribution
Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada
Recall date
May 16, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls