FDAClass IIJune 7, 2023

QUADROX-i Adult recall: Sterility Concern

QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item no. HMO 70000-USA 701067820 HMO 71000-USA 701067823

Why was QUADROX-i Adult recalled?

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Item no. UDI-DI HMO 70000-USA 701067820 4058863019147 HMO 71000-USA 701067823 4058863017341 All lots

Product
QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Model Item no. HMO 70000-USA 701067820 HMO 71000-USA 701067823
Recalling firm
Maquet Medical Systems USA
Quantity
0 (US)
Distribution
Domestic distribution nationwide.
Recall date
June 7, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls