Stryker Orthopaedics Triathlon femoral component. recall: Sterility Concern
Stryker Orthopaedics Triathlon femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total…
Why was Stryker Orthopaedics Triathlon femoral component. recalled?
The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Item number Lot number 5512-F-201 Lots JYET, KFVZ, KGLX ; 5512-F-202 Lots IPPOP, JYKX, KHZL, KJDN, KJHI, KJKY; 5512-F-301 Lots JTJA, KKAE, KKNB; 5512-F-302 Lots JXXF, KEYD, KFBY, KFTZ, KFXT, KHAB; 5512-F-401 Lots JWGY, KIED; 5512-F-402 Lots IUAJ, JJHT, JOMB, JWPF, JXXM, KGPE, KJMO, KLUP; 5512-F-501 Lots JXMV, KGKP; 5512-F-502 Lots KHWY, KJBK; 5512-F-601 Lots JIML, JIRX, KFAS, KIIY; 5512-F-602 Lots KEXE, KFOS, KFPI, KJAG, KKJW; 5512-F-701 Lot JKTF; 5512-F-801 Lots JWNL, JWPY; 5515-F-301 Lots IVXZD, JEPOD, JHZRA, JITNA, JITND,JJAFD, JJNYA, JJNYD, JKEND, JSNRD, JXBXA, JYFTD, JYGID, JYMTD,JYPJD, KFWHA, KFXID, KGAXD, KGJTD, KGXNA, KIBTA, KIEBA, KIGRD, KIYUD, KJABD, KJISD, KJRID; 5515-F-302 Lots IOLSD, IOLTD, IPRVD, IPTFD, IVGVA, IVLVD, JGIXD, JIGSD, JIYVD, JJSYA, JJWNA, JNVTA, JOWJD, JPFDA, JPHRA, JPIWA, JSZGD, JVYDD, JWPRA, JXBKD, JXOBA, JXXZD, JYBRA, JYBRD, JYKFD, JYKTA, JYMXA, JYNIA, JYNID, JYUED, JZDWD, JZEHD, JZFDD, KFNEA, KFOBA, KFOBD, KFOHD, KFVND, KGMFA, KGMFD, KGWGA, KGZFA, KHBGA, KIEXD, KJBFD, KJDLD, KJFXA, KJFXD, KJNPA, KKJOD; 5515-F-401 Lots JPDRA, JPIXD, JVLND, JVMTD, JXUDD, JYIDD, JYMUD, JZDSA, KABYA, KFDKD, KFMLA, KFPNA, KGAVD, KHEAA, KHKSA, KHZHA, KIATA, KIBVD, KIFMD, KIFOA, KIIFA, KJKBD, KKFYD; 5515-F-402 Lots JIFRD, JIGXD, JIMUD, JIRZD, JKESD, JLVSA, JLWTA, JOUGD, JPFEA, JPGKD, JYEIA, KAJID, KFSND, KGFLD, KGRBA, KGSHD, KGYOD, KJXDD, KKVHA, KLBHD, KLLBD; 5515-F-501 Lots HWHLA, IEXUA, IFDEA, IKEMD, JIAID, JILXA, JJBFD, JJTVD, JKGXD, JOXUA, JPMVD, JZRRA, KFSRA, KGIVD, KGKZA, KGSID, KGVHD, KGVPD, KHGPA, KHKAD, KHWRD, KHXWA, KJGBA, KKHFA, KKHFD, KLOSA; 5515-F-502 Lots IKRKA, JIDLA, JIEZD, JIVTA,JIYYD, JJWSA, JKAAD, JNVJA, JOPDD, JPFBD, JVUAD, JXBYD, JXFLA, JXTRD, JXYSA, JZHAD, JZKTD, KFNGD, KGTLA, KGTMA, KGYFA, KHGVA, KHGVA, KHGVD, KHOFD, KHPLA, KHRTA, KHWMA, KHWVA, KIDPA, KIJDA, KJGLD, KKMDD, KKOOD, KKPKA, KKWAA, KLBZA, KLMBA, KLSFA, KLSFD; 5515-F-601 Lots IPFTA, JIUMD, JJHZA, JMBTA, JWFHD, JWKTA, JXWXA, KABRD, KAGLA, KFISA, KFJXA, KGBVA, KGTDA, KHFSA, KHGNA, KHJAA, KHLIA, KHSWA, KHUAA, KHYED, KHZYA, KIGND, KJKOA; 5515-F-602 Lots JIYHD, JIZOA, JJBWD, JJWPD, JKDPD, JKERA, JPPKD, JJPSXD, JRBRA, JYTBD, KEZND, KFKDA, KFTMD, KFURA, KGIRA, KGNPD, KGRYD, KHJTD, KHPZA, KKXKA, KLKSA.
- Product
- Stryker Orthopaedics Triathlon femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total…
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- 1,147 in total
- Distribution
- Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.
- Recall date
- May 27, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)