Dimension clinical chemistry system, Sirolimus (SIRO) Flex reagent cartridge (DF306 SMN 10464331) Lot EB6064, Open Well Instability The SIRO method is…
Why was Dimension clinical chemistry system recalled?
Siemens Healthcare Diagnostics has received customer complaints regarding QC shifts and imprecision after one day when using Dimension¿ SIRO lot EB6064. They have confirmed SIRO lot # EB6064 does not meet the open well stability claim of 2 days.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot # EB6064
- Product
- Dimension clinical chemistry system, Sirolimus (SIRO) Flex reagent cartridge (DF306 SMN 10464331) Lot EB6064, Open Well Instability The SIRO method is…
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 394
- Distribution
- Worldwide distribution and US including Puerto Rico and to the states of : TX, FL, CA, LA, DC, GA, IL, AR, WI, MI and MD.
- Recall date
- May 18, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)