StealthStation Cranial Software recall: Software Issue
StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems
Why was StealthStation Cranial Software recalled?
During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/UDI-DI; Software Version: 9735585/00763000631635, 9735586/00763000631765, 9735587/00763000631826; 3.1.4
- Product
- StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems
- Recalling firm
- Medtronic Navigation, Inc.
- Quantity
- 746
- Distribution
- Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, Bosnia and Herzegovina, BULGARIA, CANADA, CHILE, CROATIA, CZECH REPUBLIC, DENMARK, FRANCE, GREECE, INDIA, IRELAND, ITALY, JAPAN, LUXEMBOURG, MARTINIQUE, NEW ZEALAND, POLAND, PORTUGAL, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVAKIA, SPAIN, SWITZERLAND, TAIWAN, Turkey, UNITED ARAB EMIRATES, UNITED KINGDOM
- Recall date
- June 14, 2023
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)