Component for Laser Scanning Microscopes recall
Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Why was Component for Laser Scanning Microscopes recalled?
When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
(LSM 700, LSM 800 and LSM 900)
- Product
- Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
- Recalling firm
- Zeiss, Carl Inc
- Quantity
- US: 600; Global: 1600
- Distribution
- Worldwide Distribution - US Nationwide
- Recall date
- April 15, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)