FDAClass IIMay 31, 2023

Incisive CT Plus -Whole-body computed tomography recall

Incisive CT Plus -Whole-body computed tomography (CT) X-Ray System Model Number: 728149 (OUS)

Why was Incisive CT Plus -Whole-body computed tomography recalled?

Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: (01)00884838103474(21) Serial Number: 530537 554032 554060 554065 554078

Product
Incisive CT Plus -Whole-body computed tomography (CT) X-Ray System Model Number: 728149 (OUS)
Recalling firm
PHILIPS HEADQUARTERS CAMBRIDGE
Quantity
5 units
Distribution
AR, CA, OK, TX Foreign: Country Argentina Australia Austria China Colombia Czech Republic Dominican Rep France Germany Guadeloupe Hong Kong India Japan Kuwait Latvia Lithuania Netherlands Norway Portugal Romania Spain Switzerland Taiwan Thailand United Kingdom Vietnam
Recall date
May 31, 2023
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls