NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM recall: Software Issue
NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, REF: NIM4CM01RF; PATIENT INTERFACE…
Why was NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM recalled?
Nerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead to procedure delay, extubation, and medical intervention.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All NIM Vital Nerve Monitoring Systems manufactured or installed with software version v1.6.4 or earlier. REF/UDI-DI: NIM4CM01/763000002978, 00763000395896, 00763000401597, 00763000528577; NIM4CM01RF/00763000002992; NIM4CPB1/763000002985, 00763000401603, 00763000395902, 00763000528584; NIM4CPB1RF/00763000003005; NIM4SWU143/00763000709341, 00763000869823; NIM4SWU154/00763000945398; NIM4SWU164/00763000974312
- Product
- NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0, REF: NIM4CM01; CONSOLE NIM4CM01RF NIM 4.0 REFURBISHED, REF: NIM4CM01RF; PATIENT INTERFACE…
- Recalling firm
- Medtronic Xomed, Inc.
- Quantity
- 9,788
- Distribution
- Worldwide - US Nationwide distribution including in the states of MD, MN, PA, OK, OH, FL, CO, GA, NC, KS, IL, WY, AK, NE, VA, MI, CA, NY, WI, UT, AZ, AR, TX, WA, AL, NJ, SD, MS, KY, TN, MO, OR, MT, MA, CT, NM, LA, WV, NH, DC, DE, IN, NV, IA, SC, ID, ME, HI, RI, VT, ND, GU, PR and the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Canary Islands, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Namibia, Nepal, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
- Recall date
- April 30, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)