FDAClass IIMay 27, 2015

CT LightSpeed 16 and Discovery ST Pet/CT DST recall

CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state…

Why was CT LightSpeed 16 and Discovery ST Pet/CT DST recalled?

GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Mfg Lot or Serial # System ID 00000331100CN8 904223CT 00000345975CN7 409838LS16 00000337343CN8 512OCCT 00000344962CN6 INSIGHTPCT1356

Product
CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state…
Recalling firm
GE Healthcare
Quantity
4 USA
Distribution
Distributed to the states of MN, FL & TX.
Recall date
May 27, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)