The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting recall
The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting…
Why was The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting recalled?
There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model No. VH-4000 (Hemopro 2), VH-4001 (Hemopro 2 w/ Vasoshield); UDI-DI: 00607567700406 (Hemopro 2), 00607567700451 (Hemopro 2 w/ Vasoshield); Lot/Serial No. 3000436447, 3000437191, 3000439557, 3000439600, 3000439623, 3000439922, 3000440187, 3000440220, 3000440283, 3000440754, 3000440957, 3000440981, 3000440982, 3000441144, 3000441319, 3000441541, 3000441774, 3000442133, 3000442263, 3000442323, 3000442627, 3000442871, 3000442953, 3000442955, 3000443198, 3000443216, 3000443545, 3000443632, 3000443851, 3000444131, 3000444151, 3000444484, 3000445702, 3000445704, 3000445941, 3000446003, 3000446365, 3000446489, 3000446709, 3000446774, 3000447195, 3000447535, 3000447597, 3000447845, 3000448208, 3000448471, 3000448761, 3000449047, 3000449170, 3000449267, 3000450191, 3000450967.
- Product
- The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting…
- Recalling firm
- Maquet Cardiovascular, LLC
- Quantity
- 22,244 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.
- Recall date
- April 30, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)