Expression MR400 MRI Patient Monitoring System recall: Software Issue
Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product…
Why was Expression MR400 MRI Patient Monitoring System recalled?
Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00
- Product
- Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product…
- Recalling firm
- Philips North America, LLC
- Quantity
- 2214 units
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.
- Recall date
- April 8, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)