ACCU-CHEK Compact Plus Test Strips For use with recall: Labeling Error
ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip…
Why was ACCU-CHEK Compact Plus Test Strips For use with recalled?
Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
5599423160, 5599415160, 5919967160, 5919959160, 5919983160, 5907675001
- Product
- ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip…
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- 590,136 boxes
- Distribution
- Nationwide Distribution.
- Recall date
- May 21, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)