FLS - Foot Locking Plates System and HOL - Hallux recall
FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot…
Why was FLS - Foot Locking Plates System and HOL - Hallux recalled?
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
1. Article Number (Lot Numbers): 37303-10 (18/151720, 18/268729, 679/032432, 679/072268, 679/102237, 679/102307), 37303-12 (18/570637, 679/032434, 679/072269, 679/102346, 679/112227), 37303-14 (18/751091, 26/F66913, 679/012510, 679/062288, 679/102359, 679/112410), 37303-16 (679/072263, 679/102245, 18/151841, 26/F66914, 18/331193, 18/543625, 679/012511, 679/042403, 679/102360), 37303-18 (18/151942, 18/210467, 18/231415, 18/341194, 18/572633, 18/771093, 679/012512, 679/022507, 679/042404, 679/102248, 679/112319), 37303-20 (18/152043, 18/211468, 18/351195, 18/573634, 18/701585, 18/781094, 679/022508, 679/042420, 679/112216, 679/112328), 37303-22 (18/171705, 18/212469, 18/361196, 18/574635, 18/791095, 679/022509, 679/042417, 679/112224, 679/112340), 37303-24 (18/575639, 18/741572, 26/F66915, 679/022510, 679/112226, 679/122313, 679/122333), 37303-26 (18/152346, 18/214471, 26/F66916, 679/012403, 679/112230, 679/122323), 37303-28 (26/F66917, 679/112231, 679/122328), 37303-30 (18/192716, 18/591743, 679/122334). 2. Article Number (Lot Numbers): 32271-10 (18/061949, 18/231167, 679/062351), 32271-12 (18/048927, 26/F66235, 679/042215, 679/062342, 679/062405, 679/062489, 679/072271), 32271-14 (18/049928, 18/231165, 26/F66140, 679/042210, 679/052340, 679/052518, 679/062471, 679/062490, 679/072267), 32271-16 (18/231164, 18/311125, 26/F66141, 679/042216, 679/052519, 679/062315, 679/062483, 679/072272, 679/072401), 32271-18 (18/062950, 26/F65556, 26/F66236, 679/032261, 679/062332, 679/072402), 32271-20 (18/063951, 26/F66237, 679/052326), 32271-22 (18/064952, 26/F65557, 679/032263, 679/062354), 32271-24 (18/065953, 26/F65558, 679/062333), 32271-26 (679/032265), 32271-28 (26/F65559, 679/062357, 679/052520), 32271-32 (26/F65561), 32271-34 (26/F65562). 3. Article Number (Lot Numbers): 37304-10 (18/635486, 26/F24799, 679/012419, 86/011176), 37304-12 (18/061352, 18/521172, 18/5211721, 18/576643, 18/630487, 679/012433, 679/092207, 679/122409), 37304-14 (18/081360, 18/531173, 18/5311731, 18/577644, 18/631488, 679/012434, 679/092208, 679/122410), 37304-16 (18/541174, 18/578645, 18/632489, 679/012444, 679/092201, 679/122411), 37304-18 (18/092020, 18/551175, 18/633490, 679/022403, 679/092202), 37304-20 (18/011320, 18/561176, 18/634491, 679/012287, 679/012420, 679/082217, 86/483331), 37304-22 (18/241273, 679/012288, 679/012421), 37304-24 (18/021321, 18/571177, 679/012289, 679/012445). 4. Article Number (Lot Numbers): 21413-6 (26/F52124). 5. Article Number (Lot Numbers): 21431-ME (679/102045). 6. Article Number (Lot Numbers): 21482 (679/092020), 21483 (26/F55902). 7. Article Number (Lot Numbers): 21421-5 (26/F65377, 679/092039). 8. Article Number (Lot Numbers): 37241-10 (18/012169, 18/610104, 18/701013), 37241-12 (18/231295, 18/767108), 37241-14 (18/511016, 18/537392, 26/F24807, 679/012280), 37241-16 (18/005340, 18/121106, 18/231387, 18/521017, 18/560517), 37241-18 (18/232328, 18/531018, 679/072408, 86/311848), 37241-20 (18/233388, 18/541020, 18/562519, 679/072419, 86/287847), 37241-22 (18/066109, 18/234389, 18/551021, 26/F24594, 26/F24811), 37241-24 (18/561022, 18/564521, 18/608110, 26/F24595), 37241-26 (18/611111, 18/711023, 26/F24596, 26/F24813), 37241-28 (18/612112, 18/721024, 18/811597, 26/F24597, 86/485508), 37241-30 (18/015172, 18/613113, 18/731019, 26/F24598). 9. Article Number (Lot Numbers): 21015-08 (00/0123-12, 00/0321-43, 00/0723-07), 21015-09 (00/0123-13).
- Product
- FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot…
- Recalling firm
- I.T.S. GmbH
- Quantity
- 15,517 units
- Distribution
- US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
- Recall date
- April 1, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)