MultiDiagnost Eleva 708036 As a multifunctional recall
MultiDiagnost Eleva 708036 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be…
Why was MultiDiagnost Eleva 708036 As a multifunctional recalled?
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
System Serial # 18 42 46 41 36 35 37 47 27 43 10 5 24 29 34 49 26 13 15 25 50 12 17 48 32 30 31 14 40 19 9 16 45 8 4 21 11
- Product
- MultiDiagnost Eleva 708036 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be…
- Recalling firm
- Philips Medical Systems Nederlands
- Quantity
- 37
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
- Recall date
- May 9, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)