FDAClass IIApril 8, 2020

Berchtold Chromophare Ceiling Mounted Surgical Light recall

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

Why was Berchtold Chromophare Ceiling Mounted Surgical Light recalled?

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number Range: 091316 70016 through 083117-73252

Product
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Recalling firm
Stryker Communications
Quantity
1415 Surgical Lights
Distribution
US: Nationwide OUS: Canada
Recall date
April 8, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)