FDAClass IIApril 24, 2024

SW Bari Lift & Transfer recall

SW Bari Lift & Transfer, Model Number 38060000

Why was SW Bari Lift & Transfer recalled?

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/DI 00845699001040, All serial numbers

Product
SW Bari Lift & Transfer, Model Number 38060000
Recalling firm
Raye's Inc.
Quantity
1,149 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Recall date
April 24, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls