FDAClass IIMay 26, 2021

Elecsys Troponin I STAT Immunoassay recall

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Why was Elecsys Troponin I STAT Immunoassay recalled?

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 04015630939381

Product
Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
10, 489 kits
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.
Recall date
May 26, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls