Imactis CT-Navigation System recall
Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
Why was Imactis CT-Navigation System recalled?
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
HIBC: +B681J02000; E85101LA: IMB-22-013 IMB-24-013 IMB-23-006 IMB-24-001 IMB-24-016 IMB-24-018 IMB-24-007 IMB-21-001 IMB-23-004 IMB-24-015 IMB-22-014 IMB-22-011 IMB-21-008 IMB-22-010 IMB-23-003 IMB-22-002 IMB-16-005 IMB-16-002 IMB-17-007 IMB-21-003 IMB-16-003 IMB-18-006 IMB-16-006 IMB-19-014 IMB-23-007 IMB-19-011 IMB-24-011 IMB-15-001 IMB-24-009 IMB-20-005 IMB-19-009 IMB-15-003 IMB-18-007 IMB-17-011 IMB-18-003 IMB-19-003 IMB-17-006 IMB-20-004 IMB-21-005 IMB-21-006 IMB-20-012 IMB-22-004 IMB-17-008 IMB-20-022 IMB-19-019 IMB-18-004 IMB-17-012 IMB-19-008 IMB-15-004 IMB-16-004 IMB-17-005 IMB-18-002 IMB-18-005 IMB-18-010 IMB-19-005 IMB-15-002 IMB-17-009 IMB-17-010 IMB-24-012 IMB-17-002 IMB-20-016 IMB-22-007 IMB-24-002 IMB-20-010 IMB-22-005 IMB-22-017 IMB-24-021 IMB-19-002 IMB-24-017 IMB-20-021 IMB-19-007 IMB-24-014 IMB-24-004 IMB-24-020 IMB-19-013 IMB-20-009 IMB-24-006 IMB-20-008 IMB-24-023 IMB-24-022 IMB-18-009 IMB-19-001 IMB-19-004 IMB-19-006 IMB-19-010 IMB-19-016 IMB-22-001 IMB-22-012 IMB-22-016 IMB-22-018 IMB-17-003 IMB-24-005 IMB-16-007 IMB-22-015 IMB-24-003 IMB-20-002 IMB-20-003 IMB-20-017 IMB-20-018 IMB-20-019 IMB-21-002 IMB-21-004 IMB-22-003 IMB-22-008 IMB-22-009 IMB-20-011 IMB-20-020; E85101LP: IMB-25-005 IMB-21-007 IMB-25-007 IMB-25-008 IMB-20-001 IMB-20-013 IMB-25-012 IMB-25-009 IMB-25-006 IMB-21-009 IMB-25-013 IMB-25-010 IMB-19-012 IMB-25-001 IMB-25-004 IMB-24-024 IMB-25-023 IMB-25-003
- Product
- Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 125 units
- Distribution
- Worldwide distribution - US Nationwide.
- Recall date
- April 1, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)