FDAClass IIMay 9, 2018

MHI-TM2000 Linear Accelerator System Product Usage recall: Software Issue

MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and…

Why was MHI-TM2000 Linear Accelerator System Product Usage recalled?

Due to a system controller software anomaly, the patient positioning deviation correction may not be applied and may result in the wrong part of the patient being irradiated.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number: 203919

Product
MHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and…
Recalling firm
Hitachi Ltd., Medical System Operations Group
Quantity
1
Distribution
US Distribution in the state of NY.
Recall date
May 9, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls