FDAClass IIJune 5, 2019

Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA recall: Sterility Concern

Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551

Why was Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA recalled?

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Batch Numbers: 3000041542 3000046483 3000049065 3000055328 3000057682 3000060483 3000063712

Product
Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551
Recalling firm
Maquet Cardiovascular, LLC
Quantity
N/A
Distribution
Nationwide
Recall date
June 5, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls