FDAClass IIMarch 25, 2026

MICS3 Angled Sagittal Saw Attachment recall

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Why was MICS3 Angled Sagittal Saw Attachment recalled?

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;

Product
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Recalling firm
Howmedica Osteonics Corp.
Quantity
1098 units
Distribution
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Recall date
March 25, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)