enGen (TM) Laboratory Automation System configured with recall
enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Track System…
Why was enGen (TM) Laboratory Automation System configured with recalled?
The electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
US Serial Numbers: JFBX395J, GBX395J, J953252, JJXBQW4J, J3RHC23J, J86PWC5J, J5LQ6Q2J, JFCM073J, J54FC01J, J5J4B45J, J7KSC15J, J4HX7S4J, JGWXQ52J, J953221, JG7BFB5J, JHPKG25J, JCXL941J, J736BM4J, JFXCBG3J; -- Foreign Serial Numbers: J89133 (VAS), JHWJZY4J, J3J4B45J, J953244, JCT20Q3J, J16XQH4J, GWJVY4J, DWX052J, J271264, J271368, J271371, J271269, J85206, J953216, DP2LF62, J7023874, J7023847, J7025017, J7024893, J7023645, J7023435, J57663, J800103, S/N 89VQ762 (VAS), J75863, J75691, J30372, J3055503, J3055496.
- Product
- enGen (TM) Laboratory Automation System configured with Thermo Scientific Recapper Module (230V or 110V -- COMMON/USUAL NAME: enGen (TM) Track System…
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- US: 19, Foreign: 30
- Distribution
- Distributed in the states of MD, NY, ID, NC, CO, OH, PA, TN, MS, MO, CA, DE, FL, GA, AZ, NJ, and WV and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, France, India, Italy, Norway, Portugal, Saudi Arabia, Spain, Sweden, Thailand, and United Kingdom.
- Recall date
- May 4, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)