FDAClass IIAugust 31, 2022

Nobel Biocare N1 TiUltra TCC NP 3.5x13mm recall

Nobel Biocare N1 TiUltra TCC NP 3.5x13mm

Why was Nobel Biocare N1 TiUltra TCC NP 3.5x13mm recalled?

Due to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog number: 300859 Lot numbers / UDI Codes: 12167337 / (01)07332747161717(10)12167337(11)210119(17)251218 12167830 / (01)07332747161717(10)12167830(11)210125(17)251224 12168184 / (01)07332747161717(10)12168184(11)210128(17)251227

Product
Nobel Biocare N1 TiUltra TCC NP 3.5x13mm
Recalling firm
Nobel Biocare
Quantity
347 devices
Distribution
U.S Nationwide.: AZ, CA, CT, FL, HI, NC, NY, OR, PA, TX, UT, and WI O.U.S.: Austria, Belgium, Canada, Germany, Spain, France, United Kingdom, Hungary, Iceland, Italy, Netherlands, Norway, Portugal, and Sweden
Recall date
August 31, 2022
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls