eCareCoordinator allows the customer to schedule recall: Software Issue
eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional…
Why was eCareCoordinator allows the customer to schedule recalled?
The adherence check generates a software error. Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later. An error message is generated and forwarded to operations. Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement. If the pa
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
453564506091 eCareCoordinator 1.3
- Product
- eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure). For medical use by professional…
- Recalling firm
- Philips Visicu
- Quantity
- 9
- Distribution
- US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.
- Recall date
- May 4, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)