FDAClass IMarch 18, 2026

Owner's Booklets and Instructions for Use that are used recall: Labeling Error

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX…

Why was Owner's Booklets and Instructions for Use that are used recalled?

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE4011-00/87701426520, RE4011-01/87701426254, RE4019-01/713733069816, RE4023-01/21292007256, RE4025-01/93351018992, RE4027-01/21292007065, RE4031-01/41220001286, RE4051-43/20612479197231, RE4054-00/10939722447, RE4054-01/10939719447, RE4060-01/321130788669, RE4061-01/41520147745, RE4066-01/11822445610, RE4078-40/21292007614, RE4081-03/21292014797, RE4087-01/77890394113, RE4089-00/52569138557, RE4089-01/52569138526, RE4095-40/21292007041, RE4097-01/725439988933, RE4098-01/21292008901, RE4099-40/21292014902, RE4099-45/21292008697, RE4202-43/21292007454, RE4203-01/41260016028, RE4209-01/36800234529, RE4211-01/10939958693, RE4H01-00/21292006082, RE4H01-01/21292006051, RE4H01-04/21292008895, RE4H01-40/21292006068, RE4H01-43/21292006075, RE4H01-51/21292014940. International: RE4i01-00/N/A, RE4i01-01/N/A, RE4i01-02/N/A, RE4i03-02/N/A, RE4i03-11/N/A, RE4i05-00/N/A, RE4i07-00/N/A, RE4i23-00/21292010133, RE4i29-11/N/A, RE4i61-02/N/A, RE4i61-02THI/N/A, RE4i82-11/21292012205, RE4i82-12/21292009977. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX Owner's Booklet/ RE4TVH03 Rev 56 and prior, TRUE METRIX Owner's Booklet (English)/ RE4TVH35 Rev 51 and prior, TRUE METRIX Owner's Booklet (Spanish)/ RE4TVH35S Rev 51 and prior, McKesson Med Surg TRUE METRIX Self Monitoring Owners Booklet/ RE4SUN03 Rev 55 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior, TRUE METRIX Blood Glucose Monitoring System IFU Trividia House Brand/ RE4ITV16 Rev 53 and prior, TRUE METRIX System Instructions for Use (IFU) for Trividia Health UK Limited/RE4UKT16 Rev 55 and prior, Trividia Health LATAM - TRUE METRIX System IFU mg/dL/ RE4LAT16 Rev 50 and prior, TRUE METRIX System Instructions for Use (IFU) for Trividia Health Australia/ RE4AUT16 Rev 56 and prior. Device serial numbers: Prefix "T" followed by an eight-digit numeric sequence.

Product
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX…
Recalling firm
Trividia Health, Inc.
Quantity
12,067,458
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
Recall date
March 18, 2026
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls