Link Sled Knee System Porex Endo recall
LINK SLED Knee System POREX ENDO MODEL SLED TIBIAL COMP 11 MM X 45 MM Item Number: 15-2230/03
Why was Link Sled Knee System Porex Endo recalled?
Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: 04026575429134 All lot numbers within the labeled expiration date
- Product
- LINK SLED Knee System POREX ENDO MODEL SLED TIBIAL COMP 11 MM X 45 MM Item Number: 15-2230/03
- Recalling firm
- Linkbio Corp.
- Quantity
- 5 units
- Distribution
- NY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland, Slovakia, Spain, United Kingdom, Germany, Austria, Argentina, Australia, China, India, Iraq, Jordan, Colombia, Mexico, Saudi Arabia, Ukraine, Vietnam, and Belarus
- Recall date
- May 10, 2023
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)