FDAClass IIApril 16, 2025

IVX Station Reconstitution Transfer Set recall: Sterility Concern

IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf…

Why was IVX Station Reconstitution Transfer Set recalled?

IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model number: 298921215. UDI Number: 0086000726042101124301840119041525, 0086000726042101124300840119041525, 0086000726042101124301680119041525. Lot number 011243.

Product
IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf…
Recalling firm
Omnicell, Inc.
Quantity
3
Distribution
Distributed in the US to Alabama
Recall date
April 16, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls