Dilator - Coons Taper - Used for dilating puncture recall: Sterility Concern
Dilator - Coons Taper - Used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular…
Why was Dilator - Coons Taper - Used for dilating puncture recalled?
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Ref. Number/UDI-DI/Lot Numbers: 1) JCD10.0-38-20- COONS (01)00827002039285(17) 15787236; 2) JCD12.0-38-20- COONS (01)00827002039292(17) 15774752, 15774753; 3) JCD14.0-38-20- COONS (01)00827002039322(17) 15797981, 15784149, 15797975 15797976, 15797982; 4) JCD16.0-38-20- COONS (01)00827002039476(17) 15794190, 15794191, 15808405; 5) JCD18.0-38-20- COONS (01)00827002040519(17) 15794189, 15797973; 6) JCD20.0-38-20- COONS (01)00827002042643(17) 15784152, 15784156, 15784158 15784164, 15794195, 15804033; 7) JCD 38-20- COONS (01)00827002044432(17) 15767086, 15784163, 15794203, 15806060; 8) JCD6.0-38-20- COONS (01)00827002042216(17) 15787237, 15787244 9) JCD8.0-38-20- COONS (01)00827002039278(17) 15797985, 15797999, 15798003
- Product
- Dilator - Coons Taper - Used for dilating puncture sites or catheter tracts for percutaneous placement of devices for vascular and non-vascular…
- Recalling firm
- Cook Incorporated
- Quantity
- 4,856 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AT, AU, BR, CA, CH, DE, ES, GB, HK, IE, IL, IN, IT, NZ, SG.
- Recall date
- April 24, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)