FDAClass IIMay 14, 2014

Inomax Dsir recall: Software Issue

INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for…

Why was Inomax Dsir recalled?

An anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated with low or no gas flow through the injector module, an intended NO cell failure alarm will occur. In addition, an unintended NO2 cell failure alarm may also occur. The NO/NO2 cell failure alarms are caused by an average NO/NO2 circuit concentration greater tha

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: DS20080152, DS20090218, DS20090243, DS20090271, DS20090523, DS20090598, DS20090821, DS20090925, DS20101022, DS20101076, DS20101147, DS20101374, DS20101413, DS20101420, DS20101499, DS20101637, DS20111199, DS20120099, DS20120150, DS20120169, DS20120288, DS20120306.

Product
INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for…
Recalling firm
INO Therapeutics (dba Ikaria)
Quantity
22
Distribution
Nationwide Distribution including DC and VA.
Recall date
May 14, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls