FDAClass IAugust 31, 2022

HAMILTON-C6 Intensive Care Ventilator recall

HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Why was HAMILTON-C6 Intensive Care Ventilator recalled?

Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 07630002808590, Serial Numbers: 5663, 6995, 6907, 6952, 7177, 7772, 6932, 6157, 8658, 7307, 5939, 6858, 6231, 8620, 6349, 8356, 6480, 8322, 8306, 9100, 9101, 9099, 9098, 6219, 7894, 7861, 8003, 6723, 7802, 4833, 7265, 7980, 7666, 7382, 6774, 7358, 7719, 6896, 6925, 9073, 9062, 9298, 9294, 9301, 9292, 9293, 9290, 9300, 9299, 9291, 9231, 9132, 9108, 8310, 8604, 8255, 8599, 7156, 8313, 8344, 8590, 8087, 8348, 8490, 8151, 7174, 8594, 8477, 8122, 8063, 9353, 9230, 9329, 9243, 9235, 9236, 9337, 9228, 9224, 9229, 9225, 9103, 8413, 6149, 8178, 8401, 7827, 8406, 8409, 7826, 6416, 7831, 6280, 9128, 9260, 9257, 9125, 9120, 9259, 9258, 9354, 8325, 8297, 7777, 6212, 6516, 6871, 8099, 6562, 8258, 5679, 5681, 7808, 7115, 8601, 7349, 7130, 6658, 7581, 7819, 8346, 6537, 8235, 7410, 6868, 8049, 8281, 7966, 7787, 9241, 9242, 9054, 9075, 9131, 9121, 9137, 9134, 9133, 9118, 6350, 7122, 7117, 6790, 7614, 8121, 7884, 8102, 8720, 8715, 8726, 8718, 8719, 9248, 9244, 9127, 9138, 9126, 8256, 8371, 9107, 9109, 9036, 9110, 9069, 9063, 9067, 8097, 9129, 9130, 9459, 8723, 8722, 8713, 8702, 8725, 8714, 8727, 8706, 8517, 8717, 7822, 8148, 7695, 8261, 7165, 6944, 7420, 6937, 7150, 6953, 7201, 6965, 7004, 7196, 8849, 8597, 8850, 9106, 9192, 9239, 9237, 9233, 8311, 7898, 7893, 6641, 6505, 6075, 7986, 8005, 7888, 7887, 9535, 9579, 9508, 9576, 9553, 9577, 9578, 9545, 9582, 9590, 9575, 7845, 7740, 8082, 8070, 6671, 7608, 6566, 8535, 8616, 7821, 6579, 7586, 7994, 7066, 7138, 9102, 9105, 14784, 8603, 8846, 8844, 8847, 8845, 8115, 7113, 7809, 7938, 7993, 7252, 8568, 7754, 7033, 8052, 6829, 8249, 9077, 9080, 9076, 9051, 9064, 9081, 9050, 9071, 9091, 7836, 7656, 9124, 9462, 9297, 9325, 9303, 8866, 9348, 9352, 9324, 9227, 6093, 6159, 6274, 6276, 6309, 6320, 6327, 6464, 6512, 6545, 6674, 6789, 7136, 7173, 7191, 7273, 7279, 7422, 7432, 7569, 7572, 7587, 7644, 7676, 7782, 7783, 7784, 7786, 7788, 7789, 7790, 7792, 7793, 7794, 7795, 7796, 7797, 7798, 7799, 7801, 7806, 7807, 7812, 7813, 7837, 7901, 7923, 7969, 8182, 8343, 8394, 8395, 8399, 8420, 8421, 8422, 8423, 8424, 8497, 8519, 8593, 8661, 8668, 8670, 8673, 8675, 8676, 8678, 8707, 8709, 8712, 9028, 9029, 9032, 9033, 9042, 9043, 9044, 9046

Product
HAMILTON-C6 Intensive Care Ventilator, REF: 160021
Recalling firm
Hamilton Medical AG
Quantity
358
Distribution
US nationwide distribution including Puerto Rico.
Recall date
August 31, 2022
Classification
Class I
Status
Ongoing
Agency
FDA (device)