Implantable Pulse Generators - Product Usage: used to recall
Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac…
Why was Implantable Pulse Generators - Product Usage: used to recalled?
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
all codes
- Product
- Implantable Pulse Generators - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac…
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 575815 units
- Distribution
- Worldwide distribution.
- Recall date
- May 12, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)