FDAClass IIIApril 27, 2016

Mach 4 Mouse Probe vial within the Mach 4 Universal recall: Labeling Error

Mach 4 Mouse Probe vial within the Mach 4 Universal HRP-Polymer Kit; Catalog number M4U534L; For In Vitro Use Immunology: MACH 4 Universal HRP-Polymer…

Why was Mach 4 Mouse Probe vial within the Mach 4 Universal recalled?

Biocare Medical has identified a single lot (082715) of its Mach 4 HRP Polymer kit in which some few Mach 4 Mouse Probe vials may have been mislabeled as HRP-Polymer.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot number 082715; Expiry 01/2017.

Product
Mach 4 Mouse Probe vial within the Mach 4 Universal HRP-Polymer Kit; Catalog number M4U534L; For In Vitro Use Immunology: MACH 4 Universal HRP-Polymer…
Recalling firm
Biocare Medical, LLC
Quantity
93
Distribution
US Nationwide Distribution in the states of: KS, CA, AZ, OR, FL, KS, CO, MA, KY, NY, IL, OR, MO, NC, TX, PA, OK. SC, TN, PA, TN, GA, and OR. Hungary, Canada, Boliva
Recall date
April 27, 2016
Classification
Class III
Status
Terminated
Agency
FDA (device)