FDAClass IApril 23, 2025

Tandem Mobi Insulin Pump with Interoperable Technology recall: Software Issue

Tandem Mobi Insulin Pump with Interoperable Technology

Why was Tandem Mobi Insulin Pump with Interoperable Technology recalled?

A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software versions 7.9/UDI: 00389152480114

Product
Tandem Mobi Insulin Pump with Interoperable Technology
Recalling firm
Tandem Diabetes Care, Inc.
Quantity
85 units
Distribution
US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, PA, TX, UT, VA, WA, WI OUS: N/A
Recall date
April 23, 2025
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls