FDAClass IIMay 7, 2014

The Osteo-Clage System consists of stainless steel recall

The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0…

Why was The Osteo-Clage System consists of stainless steel recalled?

Manufacturing of these devices with a grade of stainless steel that is not within specifications.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

2.0mm x 80mm SS cable/sleeve: Lot # 222078, 232175, 245032, 254368, 261903, 270659, 284155, 297281. 2.0mm SS sleeve only: Lot # 247553, 267469, 280871, 313261. 8, 7 hole SS compression plate: Lot # 300381. 10, 9 hole SS compression plate: Lot # 299641.

Product
The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0…
Recalling firm
Acumed LLC
Quantity
141 units in the US; 294 units outside the US.
Distribution
US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
Recall date
May 7, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls