IMMULITE 2000 Systems E2 Estradiol recall
IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS: L2KE22 (200 tests), L2KE26 (600 tests)…
Why was IMMULITE 2000 Systems E2 Estradiol recalled?
Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Manufacturing Lot #: All kit lots beginning 501 and above. SMN & (UDI) for L2KE22(D) US: 10702833 (00630414947907), L2KE26(D) 10702834 (00630414947914). SMN & (UDI) for L2KE22 OUS: 10381178 (00630414966106), L2KE26: 10381177 (00630414966007).
- Product
- IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS: L2KE22 (200 tests), L2KE26 (600 tests)…
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 6118 kits
- Distribution
- Worldwide distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, MI, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI. SC, TN, TX, UT, VA, WA, WI, WY, and countries of AE, AF, AL, AO, AR, AT, AU, AZ, BD, BE, BO, BG, BH, BR, CA, CH, CO, CL, CN, CO, CR, CY, DE, DO, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, IE, IN, IR, IS, IT, IQ, JO, KE, KR, LB, LT, LV, LY, ME, MK, MX, NL, NO, PA, PE, PH, PK, PL, PS, PT, PY, QA, RO, RU, SA, SE, SK, SL, SV, SY, TH, TJ, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XK, XS, ZA
- Recall date
- March 25, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)