FDAClass IIApril 29, 2015

Upper Right Telescope Unibodies recall

Upper Right Telescope Unibodies (5/pk) and Upper Left Telescope Unibodies (54/pk) A Herbst Appliance is used to affect changes in the skeletal…

Why was Upper Right Telescope Unibodies recalled?

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product No. 91695 and 91696

Product
Upper Right Telescope Unibodies (5/pk) and Upper Left Telescope Unibodies (54/pk) A Herbst Appliance is used to affect changes in the skeletal…
Recalling firm
Specialty Appliance Works, Inc.
Quantity
14,428 Universal Nuts
Distribution
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
Recall date
April 29, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls