FDAClass IIApril 17, 2024

Hillrom Welch Allyn Connex ProBP 3400 Digital Blood recall

Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed under the following models: a) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,EU…

Why was Hillrom Welch Allyn Connex ProBP 3400 Digital Blood recalled?

Baxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a) 34BFST-2, UDI/DI 732094009385; b) 34BFST-4, UDI/DI 732094009378; c) 34BFST-6, UDI/DI 732094009361; d) 34BXST-2, UDI/DI 732094009248; e) 34BXST-4, UDI/DI 732094167733; f) 34BXST-7, UDI/DI 732094009231; g) 34XFST-2, UDI/DI 732094009057; h) 34XFST-4, UDI/DI 732094009040; i) 34XFST-6, UDI/DI 732094009033; j) 34XFST-7, UDI/DI 732094009026; k) 34XXST-2, UDI/DI 732094008852; l) 34XXST-4, UDI/DI 732094008845; m) 34XXST-6, UDI/DI 732094008838; n) 34XXST-7, UDI/DI 732094008821; o) PWCD-5WT-2, UDI/DI 732094072402; p) PWCD-5WT-4, UDI/DI 732094155808; q) PWCD-5WT-6, UDI/DI 732094072389; r) PWCD-5WT-7, UDI/DI 732094072365; s) PWCD-6WT-2, UDI/DI 732094347494; t) PWCD-6WT-6, UDI/DI 732094347470; u) PWCD-6WT-7, UDI/DI 732094347463

Product
Hillrom Welch Allyn Connex ProBP 3400 Digital Blood Pressure Device, marketed under the following models: a) PROBP 3400, SUREBP, BLUETOOTH, MOBILE,EU…
Recalling firm
Baxter Healthcare Corporation
Quantity
11154 units
Distribution
Global, US: NY, IL
Recall date
April 17, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls