Stryker recall
Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was…
Why was Stryker recalled?
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02
- Product
- Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was…
- Recalling firm
- Stryker Corporation
- Distribution
- OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea
- Recall date
- March 25, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)