FDAClass IApril 24, 2024

Trilogy Evo Universal Ventilator recall

Trilogy Evo Universal Ventilator, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.

Why was Trilogy Evo Universal Ventilator recalled?

Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model No. DS2000X11B; UDI-DI: 606959052000; All units except those with software version 1.05.06.00.

Product
Trilogy Evo Universal Ventilator, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.
Recalling firm
Philips Respironics, Inc.
Quantity
9,999 units
Distribution
Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.
Recall date
April 24, 2024
Classification
Class I
Status
Ongoing
Agency
FDA (device)