Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The recall
Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated…
Why was Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The recalled?
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
See Attached Spreadsheat
- Product
- Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated…
- Recalling firm
- Ohmeda Medical
- Quantity
- 5,993 (5,080 US; 913 OUS)
- Distribution
- Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
- Recall date
- March 29, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)