CROSSCHECK(R) Plating system recall
CROSSCHECK(R) Plating system, Utility Plate, 7-Hole, REF CCP-UTN7. Orthopedic use.
Why was CROSSCHECK(R) Plating system recalled?
The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 3504937, 35049-37A, 35049-37A-1, 35049-38, 362806
- Product
- CROSSCHECK(R) Plating system, Utility Plate, 7-Hole, REF CCP-UTN7. Orthopedic use.
- Recalling firm
- Wright Medical Technology, Inc.
- Quantity
- 151 units
- Distribution
- Nationwide Distribution.
- Recall date
- April 29, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)