RayStation Software Version Numbers version numbers 8B recall
RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
Why was RayStation Software Version Numbers version numbers 8B recalled?
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Version Number (Build Number/UDI): 8B (8.1.0.47/07350002010129), 8B SP1 (8.1.1.8/07350002010204), 8B SP2 (8.1.2.5/07350002010235), 9A (9.0.0.113/07350002010174), 9A SP1 (9.0.1.142/07350002010488), 9B (9.1.0.933/07350002010266), 9B SP1 (9.2.0.483/07350002010297), 10A (10.0.0.1154/07350002010303), 10A SP1 (10.0.1.52/07350002010365), 10A SP2 (10.0.2.10/07350002010655), 10B (10.1.0.613/07350002010310), 10B SP1 (10.1.1.54/07350002010471), 11A (11.0.0.951/07350002010389), 11A SP1 (11.0.1.29/07350002010433), 11A SP2 (11.0.3.116/07350002010440), 11A SP3 (11.0.4.15/07350002010631), 11B (12.0.0.932/07350002010426), 11B SP1 (12.1.0.1221/07350002010495), 11B SPC1 (12.1.1.41/07350002010587), 11B SP2 (12.0.3.68/07350002010501), 11B SPC2 (12.1.2.91/07350002010617), 11B SP3 (12.0.4.12/07350002010600), 11B SPC3 (12.1.3.162/07350002010662), 11B SPC4 (12.1.4.113/07350002010709), 11B SPC5 (12.1.5.60/07350002010747), 11B SPT1 (12.3.0.119/07350002010570), 12A (13.0.0.1547/07350002010549), 12A SP1 (13.1.0.144/07350002010679), 12A SP2 (13.1.1.89/07350002010730), 2023B (14.0.0.3338/07350002010556), 2024A (15.0.0.430/07350002010723),
- Product
- RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Quantity
- 154 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hong Kong S.A.R., Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, P.R. China, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom.
- Recall date
- April 17, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)