FDAClass IIApril 9, 2025

Nasopore FD fragmentable nasal dressing recall: Sterility Concern

Nasopore FD fragmentable nasal dressing, Catalog Numbers 5400-020-108 and 5400-020-108ITL

Why was Nasopore FD fragmentable nasal dressing recalled?

There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

5400-020-108 UDI-DI: 07613327117226, Lot Numbers: 2023011317 2023021717 2023030932 2023040317 2023083017 2023102517 2024020517 2024080717; 5400-020-108ITL UDI-DI: 07613327117226, Lot Numbers: 2023012017 2023032817 2023042017 2023091117 2023111417 2023120517 2024011817 2023020732 2023041117 2023062017 2023092017 2023112017 2024010817 2024012417 2023031717 2023041417 2023082317 2023103117 2023112817 2024011217

Product
Nasopore FD fragmentable nasal dressing, Catalog Numbers 5400-020-108 and 5400-020-108ITL
Recalling firm
Stryker Corporation
Quantity
182,344 total units
Distribution
US Nationwide distribution.
Recall date
April 9, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls