ARCHITECT STAT Myoglobin Reagent Kit recall
ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.
Why was ARCHITECT STAT Myoglobin Reagent Kit recalled?
The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
LN 2K43-25 - Lot number 50808UN23, exp. 11/30/2024, UDI (01)00380740003302 (17)241130(10)50808UN23; LN 2K43-20 - Lot number 60104UN23, exp. 11/30/2024, UDI (01)00380740003296 (17)241130(10)60104UN23.
- Product
- ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.
- Recalling firm
- Abbott Laboratories
- Quantity
- 7482 kits
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IL, KS, LA, MI, MO, MS, NE, NY, OH, OR, PA, TX, VA, and WI. The countries of Angola, Austria, Belarus, Belgium, Brazil, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Mali, Mexico, Morocco, Poland, Portugal, Peoples Republic of China, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Switzerland, Taiwan, Tajikistan, Tanzania, Trinidad & Tobago, Turkey, Uganda, Uruguay, Uzbekistan, and Vietnam.
- Recall date
- April 10, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)