FDAClass IIMarch 4, 2026

Octopus Nuvo Tissue Stabilizer recall

Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

Why was Octopus Nuvo Tissue Stabilizer recalled?

During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 00763000543693; Lot Number: 0232837110, Additional lots as of 5/22/26: 0231547054 0231591827 0231758880 0231758898 0231758932 0231792414 0231792421 0231912903 0232670901 0232709219 0232709254 0232744928 0232784925 0232784989 0232805458

Product
Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
Recalling firm
Medtronic Perfusion Systems
Quantity
939 units
Distribution
US Nationwide. Global Distribution.
Recall date
March 4, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls