FDAClass IIMarch 4, 2026

Adapt Pump recall

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX…

Why was Adapt Pump recalled?

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

Product
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX…
Recalling firm
Agiliti Health - Ellis
Quantity
4286
Distribution
US Nationwide distribution.
Recall date
March 4, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls