Instructions for Use and Patient Manual for HeartWare recall
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101…
Why was Instructions for Use and Patient Manual for HeartWare recalled?
IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All serial numbers as of February 26, 2021. Model, GTIN: 1100, No GTIN Available; 1101, No GTIN Available; 1102, No GTIN Available; 1103, 00888707000017; 1104, No GTIN Available; 1104JP, 00888707000697; 1205, No GTIN Available; MCS1705PU, 00888707005364; 1403US, 00888707000475; 1407AU, 00888707001663; 1407CA, 00888707002851; 1407CH, 00888707001694; 1407DE, 00888707001700; 1407GB, 00888707001717; 1407IL, 00888707001724; 1407IN, 00888707001731; 1407IT, 00888707001748; 1407JP, 00888707001755; 1407KR, 00888707001762; 1420, 00888707000420; 1420JP, 00888707000437; 1430AR, 00888707000826; 1430AU, 00888707000833; 1430CA, 00888707000857; 1430CH, 00888707000871; 1430DE, 00888707000864; 1430GB, 00888707000888; 1430IL, 00888707000901; 1430IN, 00888707000505; 1430IT, 00888707000895; 1430JP, 00888707000918; 1430US, 00888707000307; 1440, 00888707001885; 1575, 00888707008051; 1650, 00888707000376; 1650CA-CLIN, 00888707001588; 1650DE, 00888707001373; 1475, 00888707008037; 2050, 00888707008082; 2050IL, 00888707001410; 2050US, 00888707000048; 2060, 00888707008099; 2060IL, 00888707001427; 2060US, 00888707000055;
- Product
- Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101…
- Recalling firm
- Heartware, Inc.
- Quantity
- 23123 (11292 US)
- Distribution
- Worldwide distribution, including: US nationwide, Italy, Great Britain, Germany, Japan
- Recall date
- April 28, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)