FDAClass IIMay 2, 2018

The humeral stems trials are recall

The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral Stem Trial, 10mm TRL092002511 Humeral Stem Trial, 11mm TRL092002512 Humeral Stem Trial, 12mm TRL092002513 Humeral Stem Trial, 13mm TRL092002514 Humeral Stem Trial, 14mm TRL092002515 Humeral Stem Trial, 15mm TRL092002516 Humeral Stem Trial, 16mm

Why was The humeral stems trials are recalled?

Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in surgery, with a variance of medical intervention required.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

TRL092002506 Lot Numbers: 19419-1 15916-1 15457-1 201274-1 201893-1 2046177 TRL092002507 Lot Numbers: 19419-2 15916-2 15457-2 201274-2 201893-2 2046178 TRL092002508 Lot Numbers: 19419-3 15916-3 15457-3 201274-3 201893-3 2046179 TRL092002509 Lot Numbers: 19419-4 15916-4 15457-4 201274-4 201893-4 20461710 TRL092002510 Lot Numbers: 19419-5 15916-5 15457-5 201274-5 201893-5 20461711 TRL092002511 Lot Numbers: 19419-6 15916-6 15457-6 201274-6 201893-6 20461712 TRL092002512 Lot Numbers: 19419-7 15916-7 15457-7 201274-7 201893-7 20461713 TRL092002513 Lot Numbers: 19419-8 15916-8 15457-8 201274-8 201893-8 20461714 TRL092002514 Lot Numbers: 19419-9 15916-9 15457-9 201274-9 201893-9 20461715 TRL092002515 Lot Numbers: 19419-10 15916-10 15457-10 201274-10 201893-10 20461716 TRL092002516 Lot Numbers: 19419-11 15916-11 15457-11 201274-11 201893-11 20461717

Product
The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral Stem Trial, 10mm TRL092002511 Humeral Stem Trial, 11mm TRL092002512 Humeral Stem Trial, 12mm TRL092002513 Humeral Stem Trial, 13mm TRL092002514 Humeral Stem Trial, 14mm TRL092002515 Humeral Stem Trial, 15mm TRL092002516 Humeral Stem Trial, 16mm
Recalling firm
Integra LifeSciences Corp.
Quantity
1619 units
Distribution
Worldwide Distribution - USA (nationwide) Distribution.
Recall date
May 2, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)