FDAClass IIApril 25, 2018

Cook Vacuum Pump recall

Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte…

Why was Cook Vacuum Pump recalled?

Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All lots

Product
Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte…
Recalling firm
Cook Inc.
Quantity
2,267
Distribution
Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Recall date
April 25, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls